Orange arrow Is there an Overtemp Hardware Error (13) or (15) on the APR-5000-XL?

448 views
  Possible Causes: 1) Review profile plot. If ramp rate to overtemp does not look realistic (there is a vertical upshoot) then possible bad RTD or bad RTD connection. 2) There is a software lockup. 3) Review profile plot. If ramp rate to overtemp looks realistic, then it may be bad top heater cont...
Tagged: plot, overtemp, 13