Orange arrow Is there no airflow or reduced airflow at the Reflow Head on the APR-5000-XL?

501 views
  Possible Causes: 1) Kinked air hose or hose not connected. 2) Incorrect setting in the software. 3) Bad connection. 4) No control voltage. 5) Bad reflow pump. Possible Solutions (corresponding to the Causes #): 1) Check the white-striped hose connected to the lower inlet on the reflow heater...
Tagged: pump, airflow, hose


Orange arrow Is there no or low Pre-Heater (bottom) heat on the APR-5000-XL?

558 views
  Possible Causes: 1) There is a bad connection. 2) Incorrect off-set setting in software. 3) No or low airflow through Pre-Heater. 4) Bad heater. 5) Bad solid state relay. Possible Solutions (corresponding to the Causes #): 1) Check connections to all heaters, both solid state relays, and the...